Journal article

Comparison of the In Vitro and In Vivo Electrochemical Performance of Bionic Electrodes

AR Harris, C Newbold, D Stathopoulos, P Carter, R Cowan, GG Wallace

Micromachines | Published : 2022

Abstract

The electrochemical performance of platinum electrodes was assessed in vitro and in vivo to determine the impact of electrode implantation and the relevance of in vitro testing in predicting in vivo behaviour. A significant change in electrochemical response was seen after electrode polarisation. As a result, initial in vitro measurements were poor predictors of subsequent measurements performed in vitro or in vivo. Charge storage capacity and charge density measurements from initial voltammetric measurements were not correlated with subsequent measurements. Electrode implantation also affected the electrochemical impedance. The typically reported impedance at 1 kHz was a very poor predictor..